TIP #160 Problem with the Cassette or the FMS Fluid Management System
MESSAGE DISPLAYED
Advisory xxx: Reinsert FMS. Replace FMS if problem persists.

Click here to get information on upcoming courses
Advisory xxx: Reinsert FMS. Replace FMS if problem persists.

Most of the 160 number errors deal with the cassette mechanism. If the cassette is not seated properly, the INFINITI will sense the miss alignment and ask to have the cassette reinserted. Sometimes there is a problem with the cassette that keeps it from properly seating and it fails the cassette test. The cassette test is initiated the moment the system draws in the FMS and tries to seat the cassette and then test the vacuum seals. The solution is to reseat the cassette by removing it totally from the cassette bay and make sure the area is clear of debris and salt.
If it consistently fails, try another cassette and if that cassette fails, the mechanism has an alignment or an internal fault that will require a service call. Call MEDiSURG at 317-663-3158 and we can walk you thru additional checks to see where the problem exists. This can save a service call. If your biomeds at the hospital want training, contact us about our next service school on the ALCON INFINITI. Nothing like having a service person on site to correct minor issues.